![Acrel 300286 Medical IPS Monitoring System Insulation Monitor Aim](/uploads/s/76ab9c131c6d444cb448aafa607034e2.webp)
Acrel 300286 Medical IPS Monitoring System Insulation Monitor Aim
Product Description AIM-M100 Medical Intelligent Insulation Monitor for Medical Insulation System GeneralHealthcare OT o
Description
Basic Info.
Model NO. | AIM-M100 |
Optional Function(2) | Fault Alarm Output |
Optional Function(1) | Double Communication |
Transport Package | Fob Shanghai |
Trademark | ACREL |
Origin | China |
Production Capacity | 10000PCS/Per Month |
Product Description
Product DescriptionAIM-M100 Medical Intelligent Insulation Monitor for Medical Insulation System
GeneralHealthcare OT operating theater isolation power systems / IPS for electrical safety Medical insulation monitoring device adopts advanced microcontroller technology, which has high integration, compact size,convenient installation and integrates intelligence, digitalization and networking in one. It is ideal selection for insulation monitoring of isolation power system in Class 2 medical locations such as operating room and intensive care unit.
Features* real-time monitoring and fault alarming of the ground insulation resistance,* transformer load current and transformer winding temperature of the monitored IT system,* be used with insulation fault locator, remotely starting fault-locating and displaying locating results when there areinsulation faults,* real-time monitor the line disconnection fault, temperature sensor disconnection fault and the functional grounding linedisconnection fault of the monitored system, and give the alarm indication within 2S after the fault occurs,* relay alarm output, LED alarm indication and other faults indication functions,* used for centralized alarm and display instrument, test signal generator, insulation fault locator and upper computer managementsoftware communications, and can monitor the operation status of IT system in real time.* Events logging function, which can record alarm occurrence time and fault type and is convenient for operation personnel toanalyze the operation conditions of system and promptly eliminate the faults.
Technical Parameters
Model | AIM-M100 |
Voltage | AC220V( fluctuation range ±10%) |
Frequency | 50/60Hz |
Max power consumption | <8VA |
Insulation resistance measuring range | 10~999kΩ |
Relative percentage accuracy | 0~±10% |
Alarm value range | 50~999kΩ |
Response time | <2S |
Measuring voltage | <12V |
Measuring current | <50μA |
Measuring range | 2.1~50A |
Alarm value range | 5~50A |
Measuring accuracy | class 5 |
Display | Digital and percentage display |
thermal resistor | PT100 |
Measuring range | -50ºC~+200ºC' |
Alarm value range | 0ºC~+200ºC' |
Output type | 2 channel relay output(unable to program) |
Contact rating | AC 250V/3A DC30V/3A |
Working temperature | -10ºC~+55ºC' |
Storage temperature | -20ºC+70ºC' |
Relative humidity | 5%~95%,non-condensing |
Altitude | ≤2500m |
Certificate
Prev: PTC Ceramic Heating Element in Aluminium Heater
Next: Insulation Monitoring Device Display Unit for Medical Isolated Power Supply
Our Contact